DEPUY IRELAND VELYS SAW INTERFACE LEFT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 451570105
Device Problem Premature Separation (4045)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Concomitant medical devices and therapy dates, robotic assisted base station device and satellite station device, (b)(6) 2022.Udi: (b)(4).
 
Event Description
It was reported that during a total knee arthroplasty surgical procedure, it was observed that the robotic assisted saw interface (sasi) was accidentally hit by the surgeon where it connects the robotic assisted satellite station and base station device.It was reported that when this happened the lever popped open causing an error saw3 can318.It was reported that they transitioned to a traditional surgery without any negative impact to the patient.It was reported that there were no delays in the surgical procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW INTERFACE LEFT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15868149
MDR Text Key307605300
Report Number1818910-2022-23623
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451570105
Device Lot NumberJ44924
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/31/2022
Initial Date FDA Received11/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-