MAUDE Adverse Event Report: BIOSENSE WEBSTER INC WEBSTER ® CS CATHETER WITH EZ STEER¿ TECHNOLOGY; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC WEBSTER ® CS CATHETER WITH EZ STEER¿ TECHNOLOGY; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE Back to Search Results
Model Number BD710FJ282CT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On 21-nov-2022, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.(b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a webster ® cs catheter with ez steer¿ technology and a broken tip issue occurred.It was reported that the distal tip of the a webster ® cs catheter with ez steer¿ technology was bent, and it looked like a piece of metal was poking out of the side of the tip.Bwi representative stated this was discovered right out of the package.The catheter was replaced and the issue resolved.The damage did not result in wires being exposed and did not result in any lifted or sharp rings.The issue was found proximal to electrode 2 from the distal end of the catheter.The catheter was not pre-shaped.Damage was noticed before attempting to place in the patient¿s body.
 
Additional Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a webster ® cs catheter with ez steer¿ technology and a broken tip issue occurred.It was reported that the distal tip of the a webster ® cs catheter with ez steer¿ technology was bent, and it looked like a piece of metal was poking out of the side of the tip.Bwi representative stated this was discovered right out of the package.The catheter was replaced and the issue resolved.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.The returned device's visual inspection was performed following bwi procedures.Visual analysis was performed, and no broken tip was observed, the reported issue could be related to the anchor window, it was observed a polyurethane (pu) application between the electrodes.No exposed internal parts or malfunctions were observed.The issue reported by the customer could not be replicated during the product investigation.A manufacturing record evaluation was performed for the finished device, and no internal action was found during the review.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEBSTER ® CS CATHETER WITH EZ STEER¿ TECHNOLOGY
Common Device Name
ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key15868182
Report Number2029046-2022-02980
Device Sequence Number5218654
Product Code DRF
UDI-Device Identifier10846835002423
UDI-Public10846835002423
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K090898
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2025
Device Model NumberBD710FJ282CT
Device Catalogue NumberBD710FJ282CT
Device Lot Number30859612M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BI DIR 7FR DEF CS,F-J,12 PIN
Patient SexUnknown
-
-