Catalog Number 912031
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during the surgery the anchor pulled out when the surgeon applied tension to the suture.The tip of the inserter was fractured, and all pieces were recovered.Another device was used to complete the procedure.No harm to the patient has been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection shows item and lot number are not confirmed as they are not etched on the part.It is confirmed that the anchor appears to be out of the juggerknot assembly but not broken.It is confirmed that the tip is bent and fractured, and tip jaws (prongs) are missing.It was returned without original packaging.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.5MM #2 MB SNGL
Common Device Name
FASTENER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15868230
Report Number0001825034-2022-02673
Device Sequence Number5209542
Product Code MBI
UDI-Device Identifier00880304509627
UDI-Public(01)00880304509627(17)260924(10)0002392608
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150768
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/24/2026
Device Catalogue Number912031
Device Lot Number0002392608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer03/17/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Patient SexUnknown
-
-