| Model Number |
C311 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The last calibration performed on (b)(6) 2022 was acceptable.Provided data show that quality control recovery is not stable.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received randomly questionable results for patient urine samples tested with tpuc3 (total protein urine/csf gen.3) on a cobas c 311 analyzer.The following example was provided by the customer.The sample initially resulted in a value of 0.0 mg/dl.The sample was repeated on a different site and resulted in a value of 12 mg/dl.No questionable result was reported outside of the laboratory.The tpuc3 reagent lot was 626934 with an expiration date of 31-jul-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of the reaction monitor shows an unstable reaction.The alarm trace was requested but not provided for the day of the event.The issue is consistent with a handling error as further investigation at the customer's site indicates that someone had altered the measurement limits in the application settings.Per product labeling: "measuring range: 40-2000 mg/l (4 - 200 mg/dl)".The investigation could not identify a product problem.
|
| |
|