| Model Number |
02.12.0310FL |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Loss of Range of Motion (2032)
|
| Date of Event |
05/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 6 months after the primary, the patient came in reporting stiffness in the knee and the cause is unknown.The surgeon revised the poly and the surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 14 november 2022.Lot 2114924: (b)(4) items manufactured and released on 13-dec-2021.Expiration date: 2026-11-22.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
|
| |
|