(b)(4).Only event year known: 2022.Batch # 901a99.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs21b device arrived with no apparent damage.The breakaway washer was present, cut and there were no staples present, indicating that the device achieved a full firing stroke.The device was reloaded with staples, a new washer was placed on the device, and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: the firing stroke must be completed.Do not partially fire the instrument.Incomplete firing can result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.Ensure that the firing trigger is fully squeezed to ensure proper staple formation and cutting of tissue.The event described could not be confirmed as no malformed staples were observed and the reload performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number 901a99, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
|
It was reported that during the gastric bypass procedure, the device was fired by the doctor.After firing, the stapler would not release.The doctor than retired the device a second time and breakaway washer was heard.When stapler was released, surgeons claimed ¿the posterior part of the anastomosis held and about 90% fell apart with unformed staples.The surgeon then used a linear stapler to complete anastomosis and performed a leak test.Leak test was negative, and patient is recovering well.The surgery was delayed.Unknown of how long.The procedure was completed successfully with no patient sequences.
|