MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941828450
Medical Device Problem Codes Break (1069); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.The 100% stenosed target lesion was located in the mildly tortuous and moderately calcified right coronary artery.A 4.50 x 28mm synergy xd drug-eluting stent was advanced through a non-boston scientific guide catheter.However, during introduction, advancing difficulty was encountered and the device broke mid-shaft.The device was pulled out and the procedure was completed with another synergy xd drug-eluting stent.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd 4.50 x 28mm stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube identified a break on the hypotube shaft located approximately 44cm distal to the distal end of strain relief as well as multiple kinks.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.No issues were noted.The device could be loaded on a test guidewire and guide catheter without issues.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft break occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.The 100% stenosed target lesion was located in the mildly tortuous and moderately calcified right coronary artery.A 4.50 x 28mm synergy xd drug-eluting stent was advanced through a non-boston scientific guide catheter.However, during introduction, advancing difficulty was encountered and the device broke mid-shaft.The device was pulled out and the procedure was completed with another synergy xd drug-eluting stent.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15869211
Report Number2124215-2022-49144
Device Sequence Number17447729
Product Code NIQ
UDI-Device Identifier08714729981220
UDI-Public08714729981220
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2023
Device Model NumberH7493941828450
Device Lot Number0027635993
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER - 6F CONVEY AL1; GUIDE CATHETER - 6F CONVEY AL1
Patient SexUnknown
-
-