|
Initial reporter occupation: non-healthcare professional.It has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Patient allegedly received an implant on 27jul2013 due to deep vein thrombosis (dvt)/ pulmonary embolism (pe).Patient is alleging vena cava perforation, and tilt.Patient alleges "shortness of breath" (sob).Per ct report, "the proximal aspect/head of the filter is about the axial level of the roof of the left renal vein.The right renal vein, its mid axis is about 2 mm below that of the left.The axis of the filter when viewed coronally is about 15 [degrees] tilted to the right from superior to inferior.There is equidistant deployment of the prongs/legs of the filter both the short and the long ones.The ventral right longus strong ends 4-5 mm outside the contour of the ivc.There is no fragmentation/displacement of any of the parts of the filter.Ivc demonstrates normal caliber throughout.Intraluminal ivc patency cannot be confirmed without contrast".
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: a2, a4, b1, b5, b6, b7, d1, d4, g4, h4, h6 g4 (510k): k211874 investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava perforation, tilt, shortness of breath (sob).The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported shortness of breath (sob) is directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on the work order for the lot.No other complaints on lots.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|