EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 4 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number FEMORAL COMPONENTS
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094); Osteolysis (2377); Implant Pain (4561)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal that a patient with bilateral knee replacement states that, "the right knee was replaced first and this one failed." the patient has pain and "suffering." "the doctor wants me back annually to watch the other knee which was done last year with the same product but to date shows no issues.".
 
Additional Manufacturer Narrative
It was reported via legal that a patient with bilateral knee replacement states that, "the right knee was replaced first and this one failed." the patient has pain and "suffering." "the doctor wants me back annually to watch the other knee which was done last year with the same product but to date shows no issues." based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined, insufficient information.These devices are used for treatment not diagnosis.
 
Additional Manufacturer Narrative
1038671-2024-04478.Recall number: z-0021-2022 the following sections were updated: a2, d4, d6a/d6b, d8, d10, g3/g4, h4, h6.The following sections were corrected:.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 4 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15870208
Report Number1038671-2022-01512
Device Sequence Number6698868
Product Code JWH
UDI-Device Identifier10885862001917
UDI-Public10885862001917
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2018
Device Model NumberFEMORAL COMPONENTS
Device Catalogue Number02-012-35-4015
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer11/21/2024
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/21/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR