|
A review of the device history record has been completed.No deviations or non-conformance's noted.The event of infection is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection.Further investigation: with the information collected during the investigation, there is enough evidence to support that device lot number 3258531 was manufactured under controlled conditions in accordance with abbvie¿s procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run 30032145 is not related to any additional complaint infection record reported as of today.During the trend review of all gel infection complaints for the period of nov-2020 through oct-2022, no adverse trend was noted.Given the fact that the cause of the infection cannot be specifically associated to the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) or temporary changes were identified during the investigation associated to this lot and the complaint, no corrective action is deemed necessary at this time.
|