• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-CT SOMATOM FORCE; COMPUTED TOMOGRAPHY X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-CT SOMATOM FORCE; COMPUTED TOMOGRAPHY X-RAY SYSTEM Back to Search Results
Model Number 10742326
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2022
Event Type  malfunction  
Event Description
It was reported to siemens healthineers that a malfunction occurred with the somatom force system.During an examination with a pediatric patient, communication between the gantry and the image reconstruction system (irs) was disrupted.Siemens is unaware whether the patient was rescanned following the occurrence.No negative health consequences were reported for the patient in association with the reported event.This report has been submitted with an abundance of caution.The reported event occurred in china.
 
Manufacturer Narrative
Siemens local service engineer inspected the concerned unit and resolved the issue by resetting the communication board in the irs.The system was brought back to normal operation.Internal id #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM FORCE
Type of Device
COMPUTED TOMOGRAPHY X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemenstrasse 1 or
rittigfeld 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemenstrasse 1 or
rittigfeld 1
forchheim, 91301
GM   91301
Manufacturer Contact
anastasia sokolova
40 liberty blvd.
mc 65-1a
malvern, PA 19355
6104486478
MDR Report Key15870934
MDR Text Key304810214
Report Number3004977335-2022-51471
Device Sequence Number1
Product Code JAK
UDI-Device Identifier04056869006956
UDI-Public04056869006956
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K190578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10742326
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2022
Initial Date FDA Received11/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-