(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products & mdrs: 42504605812 femur cemented plus right size 5+, lot# 64705442, mdr: 0001822565-2022-03304; 42542006402 tibia fixed cemented right size c, lot# 64320247, mdr: 0001822565-2022-03306; 42522800414 articular surface fixed bearing constrained condylar knee rt 14 mm, lot# 64202796, mdr: 0001822565-2022-03307; 42560313514 stem extension 3mm offset splined uncemented 14mm dia +135mm, lot# 64769898, mdr: 0001822565-2022-03308; 42555803205 tibial augment cemented half block rt lateral size cd 5mm, lot# 64453271, mdr: 0001822565-2022-03309; 42555803405 tibial augment cemented half block right medial size cd 5mm, lot# 64353978, mdr: 0001822565-2022-03310.
|
|
This follow-up report is being submitted to relay additional information. h6-component code- suggested code: mechanical g4-stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Review of complaint history found no additional related issues for the reported part and lot combination(s).Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: pain was reported, taking nsaids for pain, physical therapy tolerated, knee aspirations were performed.The complaint of pain and noise could not be confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|