U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8634
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the luer lock collar of two (2) clearlink system extension sets were snapped off causing a piece of the set to became lodged inside the lumen of the central line.This issue was identified during infusion.The central line had to be replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H10: one device was received for evaluation, the other sample was not received and therefore, could not be evaluated.The sample was received without the connector.A visual inspection was done which observed a separation of the connector from the tubing.The solvent appeared to be applied in a uniform manner in the circumference of the tube.The insertion of the tubing into male luer was according to specification.Dimensional tests were performed at the tubing and was according to specification.No further evaluation could be performed due to the lack of the connector.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15870976
Report Number1416980-2022-06484
Device Sequence Number17817310
Product Code FPA
UDI-Device Identifier00085412049021
UDI-Public(01)00085412049021
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K112893
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2C8634
Device Lot NumberR22H17314
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-