• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 777628
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ureteral stent was found ruptured upon opening of the package.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.No physical sample was returned, however photo samples were provided.The first photo provided shows the proximal pigtail broken off and completely separated from the body of the stent.The second photo sample was a product identifier and only showed the product packaging.Although a specific cause cannot be determined, a potential root cause for this event could be, "inappropriate package design".As no patient involvement was reported the device was not used, however, is intended for treatment purposes.It is unknown if the device had met all relevant specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "improper handling technique can seriously weaken the stent." "exercise care.Tearing of the stent can be caused by sharp instruments." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." correction: g.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the ureteral stent was found ruptured upon opening of the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15871492
Report Number1018233-2022-09006
Device Sequence Number20505566
Product Code FAD
UDI-Device Identifier10801741014564
UDI-Public(01)10801741014564
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number777628
Device Catalogue Number777628
Device Lot NumberNGGS3232
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-