• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC XPDM THORACIC PEDICLE PRB, ST; PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC XPDM THORACIC PEDICLE PRB, ST; PROBE Back to Search Results
Model Number 279702030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in germany as follows: it was reported on (b)(6) 2022, that the tip of screwdriver defective mis sleeve has a cleaning issue.Tip of thoracic pedicle probe broke off, no fragment available threaded pin for repositioning does not hold no patient involved, all items detected by customer during cleaning process.This report is for one (1) xpdm thoracic pedicle prb, st.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: device returned.H4: manufacture date added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5: event description edited for clarity.
 
Event or Problem Description
It was reported that on november 2, 2022, the following issues were discovered: the tip of screwdriver was defective.An mis sleeve had a cleaning issue.The tip of thoracic pedicle probe broke off, no fragment available.The threaded pin for repositioning does not hold.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a review of the receiving inspection (ri) for xpdm thoracic pedicle prb, st was conducted identifying that lot number nw166384 was released in two batch.¿ batch1: lot qty of 103 units were released on 07 july 2015 with no discrepancies.Supplier : depuy : norwood medical ¿ batch2: lot qty of 46 units were released on 09 july 2015 ith no discrepancies.Supplier : depuy : norwood medical as a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the xpdm thoracic pedicle prb, st was found the tip broken.The broken fragment was not returned.No other defect was found.A dimensional inspection for the xpdm thoracic pedicle prb, st was unable to be performed due to postmanufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the xpdm thoracic pedicle prb, st would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPDM THORACIC PEDICLE PRB, ST
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15872097
Report Number1526439-2022-02027
Device Sequence Number17447873
Product Code HXB
UDI-Device Identifier10705034198439
UDI-Public(01)10705034198439
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number279702030
Device Catalogue Number279702030
Device Lot NumberNW166384
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/06/2022
01/05/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date12/28/2022
01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MIS ALIGNMENT DEVICE.; REDUCTION THREADED POST.; VIPER2 T20 HEXLOBE DRIVER SHAF.
Patient SexUnknown
-
-