| Model Number |
VCP219H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.The needle was intact at the pre-operative check.The needle broke intra-op when sewing unknown tissue.The broken tip was finally found during the same procedure.The surgery was prolonged about two hours.The patient is good now.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: g07002 - device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: what measures were taken to retrieve the broken piece? b ultrasound and ct.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? no.Was there any additional tissue damage as a result of searching for the needle piece? no.What is the surgeon's opinion of consequences to the patient? product issue.What tissue was being sutured when the event occurred? bile duct.Was additional dissection required in other organs/tissues other than the target tissue? no.Was there any change in the patient¿s post operative care due to the prolonged procedure? no.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 12/19/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h6 additional h3 investigation summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one empty open foil, one labeled winding former, a detached needle, and a suture that pertain to the product code vcp219h.In order to evaluate the conditions of the returned sample, it was observed that the needle appears to be broken at the tip area.These samples will be shipped for further analysis due to the needle breakage.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached due to the sample needs additional evaluation.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Date sent to the fda: 1/27/2023.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: the product received for analysis was identified as product code vcp219h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the suture attachment of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage and macroscopic plastic deformation observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|