| Catalog Number |
03P89-24 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Vasculitis (2004)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded a suspected discrepant pt/inr result on a 70 year old female patient withphlebite.There was no additional patient information available at the time of this report.I-stat 16-nov-2022 11h35 11h48 4.0 a, i-stat 16-nov-2022 12h28 12h33 1.7 b, lab 16-nov-2022 12h45 13h28 2.3 c.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 10-jan-2023.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Ak (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|