ABBOTT MEDICAL PORTICO; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number PRT-25
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Asystole (4442)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, a 25mm portico valves was selected for an implant.The patient has a past medical history of atrial fibrillation.The implant depth of the portico valve was non-coronary cusp (ncc) 5mm and left coronary cusp (lcc) 7mm depth.The patient developed asystole after the portico was delivered.A pacemaker was implanted due to heart block.The patient was reported to be in stable condition.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of heart block , asystole and pacemaker implantation was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the valve non-coronary implant depth was 5mm and the left coronary implant depth was 7mm, deeper than the optimal 3 mm depth.Information from the field also indicated that the patient had a history of atrial fibrillation.Based on the information received the cause of the reported incident could not be conclusively determined but could have been caused by a combination of the patient's medical history and implant depth of the valve.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Please note that per the instructions for use, "prior to release, use appropriate imaging to ensure the struts at the inflow (lvot) portion of the valve are aligned and confirm the depth of the implant is approximately 3 mm (0.12 in.).".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTICO
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15874433
Report Number2135147-2022-02240
Device Sequence Number12134230
Product Code NPT
UDI-Device Identifier05415067012487
UDI-Public05415067012487
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/21/2024
Device Model NumberPRT-25
Device Catalogue NumberPRT-25
Device Lot Number7976538
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-