• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Model Number 1506-10-006
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Dr (b)(6) revised an attune primary today for instability.He did not do the primary case and revised with zimmer so none of our products were used.Dr (b)(6) stated the index surgery likely took too much off the distal femur which led to tight flexion and loose extension gaps.He stated it was technique related not implant related.The primary procedure was done by dr (b)(6) on (b)(6) 2020 but unknown hospital.I do not have the lot numbers as we could not read them, will try to get details from customer service.
 
Event or Problem Description
Additional information received: a.Was instability related to polyethylene wear? no poly wear.No instability not due to poly wear.As stated he believed that too much distal femur was resected.B.Were any components undersized on the primary surgery that led to the instability? no malposition.The extension gap is too loose.Components sized appropriately c.Was the femur or tibia excessively resected/joint line raised during the primary surgery that led to the instability? yes as stated in the original complaint the distal femur appeared to be over resected.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 6 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15874543
Report Number1818910-2022-23723
Device Sequence Number2492289
Product Code NJL
UDI-Device Identifier10603295042600
UDI-Public10603295042600
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1506-10-006
Device Catalogue Number150610006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer12/08/2022
12/19/2022
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date12/09/2022
12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FEM RT SZ 6 NAR CEM; ATTUNE CR RP INSRT SZ 6 5MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight87 KG
-
-