| Model Number |
EQUINOXE REVERSE 46MM HUMERAL LINER +0 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
11/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: equinoxe reverse tray adapter plate tray +0 (cat# 320-10-00 / serial#: (b)(4), equinoxe reverse 46mm glenosphere (cat# 320-01-46 / serial#: (b)(4), eq reverse torque defining screw kit (cat# 320-20-00 / serial#: (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, approximately five years post initial right side tsa, the 62 y/o male patient was lifting a sign overhead and had sharp pain.Patient experienced disassociation of polyethylene.Patient had a revision approximately a month later and the event was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3 the revision reported may have been the result of incomplete seating of the liner during implantation, bone impingement, patient-related conditions, an unreported post traumatic event, or any combination of these possibilities, which led to humeral liner disassociation.However, this cannot be confirmed as the devices were not returned for evaluation and images/radiographs were unable to be obtained.The most probable root cause associated with the reported event of ¿humeral liner disassociation¿ is associated with post-operative disassociation of the humeral liner component from the humeral adapter tray/plate.Do not use for humeral liner seating difficulty.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial right reverse total shoulder arthroplasty.Subsequently, it was reported that the patient experienced pain and disassociation of the humeral liner from the humeral tray.As a result, the patient underwent a right shoulder revision surgery approximately 5 years after initial implantation.No further patient impact reported.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b3, b5, d8, d10, g1, h6 - health effect - clincal code, component code, type of investigation, investigation findings, investigation conclusions and h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h6 - type of investigation and h11.The revision reported may have been the result of incomplete seating of the liner during implantation, bone impingement, patient-related conditions, an unreported post traumatic event, or any combination of these possibilities, which led to humeral liner disassociation.However, this cannot be confirmed as the devices were not returned for evaluation and images/radiographs were unable to be obtained.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|