• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's left eye due to subjective visual disturbances.There was no vitrectomy, incision enlargement, or sutures required.The patient has fully recovered.Through follow up, it was learned that another johnson & johnson lens, model dib00 17.5 diopter was implanted as a replacement.No further information was available.
 
Additional Manufacturer Narrative
Date of event: exact date unknown/not provided.Best estimate date is between 2/8/2022-10/31/2022.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt was made to obtain missing information; however, the account did not provide the information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: feb 28, 2023.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the iol was received inside of a plastic cap.The iol was received cut in half.Visual inspection under magnification found no further issues.Based on the returned condition of the lens, no further evaluation was performed.The complaint issues; explant and dysphotopsia could not be identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15874754
Report Number3012236936-2022-02921
Device Sequence Number19462416
Product Code MFK
UDI-Device Identifier05050474750562
UDI-Public(01)05050474750562(17)240929
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/29/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU145
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date03/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-