| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm vticm5_12.6; -11.00/3.0/085 (sphere/cylinder/axis) implantable collamer lens into the patient's eye.The surgeon reports the lens was explanted and replaced with a shorter length lens due to the length of the lens.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned in microcentrifuge vial in liquid.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specification.Claim# (b)(4).
|
| |
|