Model Number 1011882
Medical Device Problem Codes Peeled/Delaminated (1454); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily tortuous lesion in the proximal right coronary artery (prca).A 014 ht bmw elite guide wire (gw) was being advanced through the guide catheter when resistance was felt between the guide wire and the guide catheter.The guide wire was removed and once outside the patient anatomy, segments of the guide wire felt and looked like it had been shaved.Imaging showed no guide wire coating in the anatomy and no foreign material was removed from the patient.A new bmw guide wire was used to successfully complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual, dimensional, and functional inspection was performed on the returned device.The reported difficult to advance was confirmed.The reported peeled/delaminated could not be confirmed; however, it is possible that the scratched teflon coating noted during return analysis may have been inadvertently noted as peeled/delaminated.The reported difficult to remove could not be tested as it was based on operational context.Additionally, it was noted that coils were misaligned, teflon was scratched and solder was separated.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.In this case, it is possible that inadvertent damage to the guide wire, such as the misaligned coils noted during return analysis, may have contributed to an interaction between the guide wire and guide catheter, resulting in the reported difficult to advance and subsequent difficult to remove.Manipulation of the guide wire during its interaction with the guide catheter may have contributed to the noted scratched teflon and material separation (solder); however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRES
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15874928
Report Number2024168-2022-11970
Device Sequence Number20205249
Product Code DQX
UDI-Device Identifier08717648144820
UDI-Public08717648144820
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K103101
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number1011882
Device Catalogue Number1011882
Device Lot Number2032861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date11/28/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale