H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultrascore 014 pta dilatation catheter was returned for evaluation.The strain relief was noted to have become loose and dislodged from its original position.The balloon noted to be twisted, no others anomalies noted during the visual evaluation.No functional testing was performed due to the nature of the complaint.Therefore, the investigation was confirmed for the reported dislodging of strain relief, as the dislodging of strain relief could be noted during the visual evaluation of the returned device.A definitive root cause for the reported dislodging of strain relief could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 04/2024), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|