• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. AHTB, ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. AHTB, ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 72203270
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
 
Event or Problem Description
It was reported that during a hip arthroscopy surgery, the bolt of the unkn ilizarov exfix instr device that clamps the traction slider in position, came apart and was not able to tighten the clamp sufficiently.The surgeon used a vice grip to resolve this problem.Once that was fixed, the orange straps that hold the foot into the support were sliding through the clasps allowing the heal to be raised out of the support.Therefore, the procedure could not be performed.The patient was already under anesthesia and had to be woken up.The procedure has been postponed.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H6: code correction.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Correction in h6 (medical device problem code).
 
Additional Manufacturer Narrative
H11: this report was inadvertently submitted under manufacturer number ¿1020279¿, the correct manufacturer number is ¿3003604053.¿.
 
Event or Problem Description
It was reported that during a hip arthroscopy surgery, the bolt of the universal hip distractor device that clamps the traction slider in position, came apart and was not able to tighten the clamp sufficiently.The surgeon used a vice grip to resolve this problem.Once that was fixed, the orange straps of the ahtb, active heel traction boot device that holds the foot into the support were sliding through the clasps allowing the heal to be raised out of the support.Therefore, the procedure could not be performed.The patient was already under anesthesia and had to be woken up.The procedure has been postponed.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is worn from use.It is attached to the hip distractor.The orange straps are very worn from use.A functional evaluation of the device found that the orange straps are able to be easily loosened by pulling due to the wear on the straps.The clamps are unable to grip fully into the straps.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states there were no clinical factors found which would have contributed to the event.The patient impact beyond the anesthesia and postponement of the procedure cannot be determined with the information provided.The root cause has been associated with component failure.Factors that could have contributed to the reported event include wear and tear from repeat use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB, ACTIVE HEEL TRACTION BOOT
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15875767
Report Number1020279-2022-04828
Device Sequence Number2518211
Product Code HST
UDI-Device Identifier00885554024821
UDI-Public00885554024821
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203270
Device Catalogue Number72202682
Device Lot NumberSN:8
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer03/06/2023
03/14/2023
03/06/2023
10/25/2023
05/22/2024
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date03/07/2023
03/15/2023
03/21/2023
10/25/2023
05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-