MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMENT, POLY/MET/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMENT, POLY/MET/POLY Back to Search Results
Model Number 71423225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 10/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery was performed on 2015, the patient experienced an aseptic loosening right total knee prosthesis.This adverse event was treated by a revision surgery on (b)(6) 2022, all implants (legion ps np fem sz 5 lt, lgn rev tibia base sz 3 rt, lgn pressfit stem 14mm x 160mm, legion con art ins 9mm sz 3-4, lgn pressfit stem 16mm x 160mm, lgn offset coupler 4mm) were exchanged.Patient's current health status is unknown.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Section h10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a right total knee arthroplasty revision was performed approximately 7 years post implantation due to aseptic mobilization (loosening).It was communicated that the requested medical documentation is not available.Without the requested clinical documentation, possible contributing clinical factors could not be definitively concluded.The patient impact beyond the reported aseptic mobilization and subsequent revision cannot be determined.No further medical assessment could be rendered at this time.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that looseness of components as a result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions, this has been identified as possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include joint laxity or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS NP FEM SZ 5 LT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMENT, POLY/MET/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15875768
Report Number1020279-2022-04833
Device Sequence Number2518212
Product Code JWH
UDI-Device Identifier03596010607607
UDI-Public03596010607607
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K060742
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71423225
Device Catalogue Number71423225
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71420523/LEGION CON ART INS 9MM SZ 3-4; 71424013/LGN REV TIBIA BASE SZ 3 RT; 71424047/LGN PRESSFIT STEM 14MM X 160MM; 71424049/LGN PRESSFIT STEM 16MM X 160MM; 71424225/LGN OFFSET COUPLER 4MM
Outcome Attributed to Adverse Event Other;
Patient Age88 YR
Patient SexUnknown
-
-