• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60AT
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during the cataract surgery with intraocular lens implant procedure, something was visually observed before implantation of the lens.After immersing the lens once in the balanced salt solution (bss) and inserting it into the patient's eye found a crack on the edge of the lens.The surgeon suspected a crack and is at the edge of the lens and there was no problem with the patient's refraction or vision.
 
Additional Manufacturer Narrative
All product and batch history records are quality reviewed prior to product release.The file indicated the use of a qualified cartridge and handpiece.The file also indicates the use of a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the ifu.The viscoelastic indicated is not qualified for the lens/company cartridge combination used.The instruction for use (ifu) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15875783
Report Number1119421-2022-02506
Device Sequence Number14073392
Product Code HQL
UDI-Device Identifier00380650130181
UDI-Public00380650130181
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/16/2025
Device Model NumberSN60AT
Device Catalogue NumberSN60AT.140
Device Lot Number15087974
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/27/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSS STERILE IRRIGATING SOLUTION.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.; UNSPECIFIED VISCOELASTIC.
Patient SexUnknown
-
-