Model Number 305616
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 8 bd sharps coll 9gal red were missing lids.The following information was provided by the initial reporter: customer received sharps containers without lids.1 case of 8, no lids on any of them.No adverse events noticed before use upon opening box.
 
Additional Manufacturer Narrative
Oem manufacture: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, (b)(4).Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that 8 bd sharps coll 9gal red were missing lids.The following information was provided by the initial reporter: customer received sharps containers without lids.1 case of 8, no lids on any of them.No adverse events noticed before use upon opening box.
 
Additional Manufacturer Narrative
H.6.Investigation summary: no product or photo was returned by the customer.It was reported by the customer that they received sharps containers without lids could not be verified due to the product not being returned for failure investigation.Based on the information provided, the device history record review process was unable to be performed since there was no lot number provided.A review of the ncmr¿s was performed; the result showed no issues reported for missing lids for the same part number throughout the last twelve months.Investigation : according with this investigation there is not enough information provided from customer such as lot number or pictures of the original packaging in order to determine the root cause of this issue.The variables that could generate this failure mode such as handling, transportation, storage or partial sales from distributor are unknown.Additional information is required to discard issue was generated by distributor facility, since partial sells and controls to handle remaining material is unknown, therefore, the likelihood of sending boxes with incorrect quantity exist.In addition, the current manufacturing controls were reviewed, and confirmed the capability to detect this type of issue (missing lids).As part of this investigation, a review of customer complaint records was performed; according to the complaint records, seven additional complaints were received through the last twelve months for the same part number and issue.These previous complaints were closed as non-manufacturing related and incomplete since there was not enough information provided from customer to determine the root cause.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.However potential root cause will be: 1.Non-controlled method to ship partial boxes to end user (re-packaging process).2.Non-controlled handling within distributor facilities.Conclusion: based on information provided it was not possible to confirm the root cause like a failure mode related to the manufacturing process since information such as method used to handle, shipped partial sells and controls to storage the remaining product within distributor facility is unknown.The controls were verified within the manufacturing process and confirmed as capable to detect the reporting failure mode (missing lids), furthermore, there were found acceptable records for every single box manufactured for this lot number within the weighing system records.If additional information that would help to determine the root cause can be provided, then a new complaint record will be open to initiate a new investigation path.All the complaint information was captured for tracking and trending purposes.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SHARPS COLL 9GAL RED
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15876448
Report Number2243072-2022-02050
Device Sequence Number20205267
Product Code FMI
UDI-Device Identifier00382903056163
UDI-Public00382903056163
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K943134
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305616
Device Catalogue Number305616
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-