• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The patient underwent percutaneous coronary intervention (pci) under local anesthesia.After implantation of coronary stent, a high-pressure balloon was planned, and a 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, after opening the package, it was noted that the shaft broke around the middle of delivery rod and could not be used normally.The physician replaced the balloon immediately to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that shaft break occurred.The patient underwent percutaneous coronary intervention (pci) under local anesthesia.After implantation of coronary stent, a high-pressure balloon was planned, and a 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, after opening the package, it was noted that the shaft broke around the middle of delivery rod and could not be used normally.The physician replaced the balloon immediately to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
(b)(6).Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The device was visually and microscopically examined.There was a separation of the hypotube at 69cm from the strain relief.The separated ends of the hypotube were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded.Product analysis confirmed the reported break as there was separation to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15876661
Report Number2124215-2022-49596
Device Sequence Number5446304
Product Code LOX
UDI-Device Identifier08714729392651
UDI-Public08714729392651
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2025
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0029608037
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/20/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-