EXACTECH, INC. NV GXL LINER LIPPED 32MM ID, GROUP 1 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NV GXL LINER LIPPED 32MM ID, GROUP 1 CUPS |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
11/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately six years post initial tha, patient was revised due to pain and wear per surgeon.Patient was revised with same sizes but with vit e poly.There was no breakage or surgery delay was noted.Patient had good range of motion (rom) and stability.Patient left in stable condition.The devices are not available for evaluation.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 4.5mm drill bit 30mm (cat#: 101-45-30 / serial#: (b)(4); biolox delta femoral head 32mm od, +3.5mm (cat#: 170-32-03 / serial#: (b)(4); alteon 6.5mm screw, 20mm (cat#: 180-65-20 / serial#: (b)(4); integrip cc, cluster 50mm, g1 (cat#: 186-01-50 / serial#: (b)(4); wedge plasma x/o sz 7 (cat#: 188-01-07 / serial#: (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the revision reported may have been the result of pain and a combination of the risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear but not limited to the combination of largest available femoral head and/or thinnest available acetabular liner.However, the photographed acetabular liner and radiograph do not show signs of wear that are inconsistent with being implanted for approximately 6 years.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected:b1, d8 e4 (remove erroneous entry from initial report), h6 clinical codes, component code, types of investigation, investigation findings and investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|