INTRINSIC THERAPEUTICS, INC BARRICAID ACD Back to Search Results
Model Number BAR-A8-10MM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Scar Tissue (2060)
Date of Event 01/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient presented with possible reherniation, and revision surgery was performed.During revision surgery, the barricaid device was found to be in proper position and performing appropriately.Device was not touched, no reherniation was found, no new defect in the disc was present.Scar tissue/adhesions were found compressing the nerve.
 
Additional Manufacturer Narrative
Based on the surgeon narrative, it was confirmed that a surgical intervention was performed to decompress a nerve root that was causing pain.During the reoperation, the surgeon confirmed that: 1) scar tissue was compressing the nerve; 2) there was no herniated nuclear material nor new annular defect present; and 3) the barricaid was still appropriately positioned and intact.No link between the scar tissue and barricaid was established.Scar tissue may develop after a discectomy regardless of whether or not a barricaid is implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRICAID ACD
Common Device Name
BARRICAID
Manufacturer (Section D)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer (Section G)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer Contact
rachael nelsen
30 commerce way
woburn, MA 01801
7819320222
MDR Report Key15878326
Report Number3006232063-2022-00002
Device Sequence Number11395941
Product Code QES
UDI-Device IdentifierM906BARA810MM0
UDI-PublicM906BARA810MM0
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P160050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/18/2023
Device Model NumberBAR-A8-10MM
Device Catalogue Number2731015-A8
Device Lot Number06152002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Initial Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-