• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6533
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A 20mm x 2.75mm nc quantum apex balloon catheter was advanced for post-dilatation.However, during the fourth inflation at 20 atmospheres for 15 seconds, the balloon ruptured.The device was removed and the procedure was completed with a 3.00mmx20mm nc quantum apex balloon catheter.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The device was visually and microscopically examined.There was contrast in the inflation lumen.There was blood and contrast in the balloon.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.At 1mm proximal from the distal markerband, there was a pinhole in the balloon.The device failed to inflate to rated burst pressure.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A 20mm x 2.75mm nc quantum apex balloon catheter was advanced for post-dilatation.However, during the fourth inflation at 20 atmospheres for 15 seconds, the balloon ruptured.The device was removed and the procedure was completed with a 3.00mmx20mm nc quantum apex balloon catheter.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15878530
Report Number2124215-2022-49177
Device Sequence Number2532736
Product Code LOX
UDI-Device Identifier08714729783473
UDI-Public08714729783473
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2023
Device Model Number6533
Device Catalogue Number6533
Device Lot Number0025947736
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-