This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the suggested event occurred due to defective charge coupled device (ccd) unit caused by the following: a) ccd cable between ccd transmission unit at distal end and connector wiring unit is broken.B) short-circuit of the ccd cable at connector wiring unit c) dielectric breakdown inside the ccd by the handling below -system ground wire was not connected.-with the system power on, remove and insert the scope connector.-dirt, moisture, and others are attached to the connector contact section.A definitive root cause cannot be identified.This information is addressed in the instructions for use (ifu): "important information ¿ please read before use: dangers, warnings, and cautions: - do not insert the video connector while the electrical contacts are wet and/or dirty.This may result in an electric shock, causing severe damage to the endoscope and compromising patient and/or operator safety.- do not coil the video cable or universal cord with a diameter of less than 12 cm.Endoscope damage can result.- turn the video system center on only when the video connector is connected to the video system center.In particular, confirm that the video system center is off before connecting or disconnecting the video connector.Failure to do so can result in equipment damage, including destruction of the image sensor." olympus will continue to monitor the field performance of this device.
|