| Model Number |
24686 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient was presented with a narrow arteriovenous fistula in the right upper limb.The target lesion was located in the stenosed basilic vein in the upper arm.A 5.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.The balloon was inflated twice at 10 atmospheres (atm) and 22 atm respectively for 30-60 seconds; however, during third inflation at 24-26 atm, the balloon ruptured.The device was removed and found to be damaged.The procedure was completed with a different device.There were no patient complications nor injuries reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated: b5: describe event or problem e1: initial reporter's address and email.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient was presented with a narrow arteriovenous fistula in the right upper limb.The target lesion was located in the stenosed basilic vein in the upper arm.A 5.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.The balloon was inflated twice at 10 atmospheres (atm) and 22 atm respectively for 30-60 seconds; however, during third inflation at 24-26 atm, the balloon ruptured.The device was removed and found to be damaged.The procedure was completed with a different device.There were no patient complications nor injuries reported and the patient status was stable.It was further reported that the target lesion was 80% stenosed, moderately calcified and mildly tortuous.
|
| |
|
Search Alerts/Recalls
|