Model Number 30975
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the string is about to rip or has pulled away.An 8 flexima apdl drainage catheter was selected for use.During preparation, it appears that the string is about to rip or has pulled away at the loop where the string is.Consequently, they did not use this particular catheter on the patient.They used another one, with the same batch number which appeared to be fine and completed the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The catheter was returned for this complaint, with its flexible cannula loaded.The suture was inspected, and no damages were observed.No damages were observed to the returned device.The flexible cannula was loaded and unloaded without problems.No issues were identified during the product analysis.
 
Event or Problem Description
It was reported that the string is about to rip or has pulled away.An 8 flexima apdl drainage catheter was selected for use.During preparation, it appears that the string is about to rip or has pulled away at the loop where the string is.Consequently, they did not use this particular catheter on the patient.They used another one, with the same batch number which appeared to be fine and completed the procedure.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA APDL
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15880172
Report Number2124215-2022-48415
Device Sequence Number2468555
Product Code FFA
UDI-Device Identifier08714729191759
UDI-Public08714729191759
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K944290
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/21/2025
Device Model Number30975
Device Catalogue Number30975
Device Lot Number0029270681
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer01/07/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-