| Model Number |
C311 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with trigl triglycerides on a cobas 4000 c (311) stand alone system.The sample initially resulted in a triglycerides value of 740.3 mg/dl and this value was reported outside of the laboratory to medical personnel treating the patient.The sample was repeated, resulting in a value of 125.8 mg/dl.The repeat value was considered correct.The triglycerides reagent lot number was 64844301, with an expiration date of 31-jul-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer confirmed that the sample initially repeated with a value of 124.6 mg/dl.A second repeat was performed after the field service engineer performed actions on the analyzer and this value was 125.8 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
An additional repeat triglycerides value was provided for the complained sample as 124.6 mg/dl.It was asked, but it is not known if this value was provided in error or if it is an additional repeat value from the sample.
|
| |
|
Additional Manufacturer Narrative
|
|
The last calibration performed on (b)(6) 2022 was ok.The provided quality control data was within range.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|