• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37601
Device Problems Failure to Deliver Energy (1211); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Seroma (2069)
Event Date 08/01/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient's ins has been working fine since 2019.The patient has been reporting since the summer of 2022 that the ins was turning off and they'd see an oor message.The impedance values were all fine.The patient was seen in november and they stated stimulation was off again however it was found to be on.It was also stated that there was fluid around the ins however the surgeon wasn't concerned.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received: it was reported that the cause was not determined.They attempted impedances with pressure and different angles but there were no changes to note.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15880503
MDR Text Key307690950
Report Number2182207-2022-02232
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2020
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Date Manufacturer Received12/06/2022
Date Device Manufactured08/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-