• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BTG ROUND LAKE VI MRI CRYOABLATION CONSOLE; UNIT, CRYOSURGICAL, ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BTG ROUND LAKE VI MRI CRYOABLATION CONSOLE; UNIT, CRYOSURGICAL, ACCESSORIES Back to Search Results
Model Number FPRCH7000
Medical Device Problem Codes Insufficient Cooling (1130); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this unit was generating suboptimal ice.A vi mri cryoablation console was used in a cryoablation procedure.The console did not generate sufficient ice.
 
Additional Manufacturer Narrative
Device eval by manufacturer: this console was serviced in the field.During inspection, gas was applied to the system and a leak was identified coming from the port area of the msp.The mts board was removed revealing that the o-ring from the gas supply line was blown out from the manifold to desiccant line on port 1.The o-ring was replaced, and functional testing passed with no further issues.A freeze/thaw cycle was completed successfully.
 
Event or Problem Description
It was reported that this unit was generating suboptimal ice.A vi mri cryoablation console was used in a cryoablation procedure.The console did not generate sufficient ice.
 
Additional Manufacturer Narrative
Device eval by manufacturer: this console was serviced in the field.During inspection, gas was applied to the system and a leak was identified coming from the port area of the msp.The mts board was removed revealing that the o-ring from the gas supply line was blown out from the manifold to desiccant line on port 1.The o-ring was replaced, and functional testing passed with no further issues.A freeze/thaw cycle was completed successfully.
 
Event or Problem Description
It was reported that this unit was generating suboptimal ice.A vi mri cryoablation console was used in a cryoablation procedure.The console did not generate sufficient ice.Additional information received indicated that after preventative maintenance was performed in the field that was thought to have resolved the issue, the console was hissing from port 3.Troubleshooting was performed by swapping sockets to another channel and checked socket torque; however, the gas was still leaking on the port 3.Technicians are scheduled to further service this console on 23jan2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VI MRI CRYOABLATION CONSOLE
Common Device Name
UNIT, CRYOSURGICAL, ACCESSORIES
Manufacturer (Section D)
BTG ROUND LAKE
4364 round lake road
arden hills MN 55112
Manufacturer (Section G)
GALIL MEDICAL
4364 round lake road
arden hills MN 55112
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15880504
Report Number2124215-2022-48574
Device Sequence Number14076742
Product Code GEH
UDI-Device Identifier00859019006318
UDI-Public00859019006318
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K173037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFPRCH7000
Device Catalogue NumberFPRCH7000
Device Lot NumberVLM003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/08/2022
01/04/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date12/21/2022
01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-