• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BTG YOKNEAM ICEPEARL 2.1 CX 90 DEGREE NEEDLE; UNIT, CRYOSURGICAL, ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BTG YOKNEAM ICEPEARL 2.1 CX 90 DEGREE NEEDLE; UNIT, CRYOSURGICAL, ACCESSORIES Back to Search Results
Model Number FPRPR3601
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle shaft was leaking gas.During device preparation for the cryoablation procedure, there was a gas bubble observed on the shaft of the icepearl 2.1 cx 90 degree needle.The device did not make contact with the patient.The needle was replaced, and the procedure was completed.There were no reported adverse consequences to the patient.
 
Event or Problem Description
It was reported that the needle shaft was leaking gas.During device preparation for the cryoablation procedure, there was a gas bubble observed on the shaft of the icepearl 2.1 cx 90 degree needle.The device did not make contact with the patient.The needle was replaced, and the procedure was completed.There were no reported adverse consequences to the patient.
 
Additional Manufacturer Narrative
Device media analysis: the device was returned for analysis.The device was hooked up to the in-house console and functional testing was performed.During initial testing, air was seen coming from the proximal end of the needle handle.Disassembly of the device confirmed a gas leak from the needle handle.Component failure of the o-ring in the handle was the cause of the handle leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICEPEARL 2.1 CX 90 DEGREE NEEDLE
Common Device Name
UNIT, CRYOSURGICAL, ACCESSORIES
Manufacturer (Section D)
BTG YOKNEAM
tavor bldg 1, industrial park
po box 224
yokneam 20692 03
IS  2069203
Manufacturer (Section G)
GALIL ISRAEL
tavor bldg 1
yoenaem 20692 03
IS   2069203
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15880591
Report Number2124215-2022-47279
Device Sequence Number5398530
Product Code GEH
UDI-Device Identifier07290104830990
UDI-Public07290104830990
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K152133
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2023
Device Model NumberFPRPR3601
Device Catalogue NumberFPRPR3601
Device Lot NumberU3539
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-