• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN039788
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Reported as "balloon did not deploy".The report states that the "iabp placed in cath lab.Alarming possible helium leak since admission with no augmenting waveform.Helium canister changed, pump exchanged, iab pulled out by [physician].The balloon was viewed and inflated manually with proximal balloon noted to be twisted.Balloon did not inflate fully.No harm or injury to the patient.No intervention necessary, patient remained stable".
 
Additional Manufacturer Narrative
Qn# (b)(4).The reported complaint of iab helium loss alarm is confirmed based on the photo provided by the customer.No part was returned to teleflex chelmsford for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the helium loss alarm.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.If the product is returned at a later date, a full investigation of the sample will be completed.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
Reported as "balloon did not deploy".The report states that the "iabp placed in cath lab.Alarming possible helium leak since admission with no augmenting waveform.Helium canister changed, pump exchanged, iab pulled out by [physician].The balloon was viewed and inflated manually with proximal balloon noted to be twisted.Balloon did not inflate fully.No harm or injury to the patient.No intervention necessary, patient remained stable".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX IAB: 7.5FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15880665
Report Number3010532612-2022-00533
Device Sequence Number2532294
Product Code DSP
UDI-Device Identifier10801902145731
UDI-Public10801902145731
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K000729
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberIPN039788
Device Catalogue NumberIAB-06840-U
Device Lot Number18F22F0019
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer12/29/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexMale
-
-