• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA PRO 2.5/15; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA PRO 2.5/15; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 393301
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
A pantera pro coronary dilatation catheter was chosen for treatment of a severely calcified lesion with 70 percent stenosis degree in the proximal lad.At the time of pre-dilation, the balloon did not inflate, it was removed and observed to be punctured.A new balloon was used for pre-dilation without complications.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned product revealed that the balloon has been partially inflated and was returned in a partially deflated state.Microscopic inspection of the guidewire exit port revealed that both the guidewire lumen and the inflation lumen are sliced open longitudinally over a length of 4.5 mm.During functional testing water was found leaking from the slit at the guidewire exit port.The balloon could only be barely inflated.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The damage at the guidewire exit port was most likely caused during the preparations for the intervention or during the insertion phase.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA PRO 2.5/15
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15880754
Report Number1028232-2022-06069
Device Sequence Number14076612
Product Code LOX
UDI-Device Identifier07640130437050
UDI-Public(01)07640130437050(17)2504
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K160985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2025
Device Model Number393301
Device Catalogue NumberSEE MODEL NO.
Device Lot Number04220791
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer01/25/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-