| Model Number |
M0035452040 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during procedure, the subject coil could not be advanced to proximal of microcatheter even after several tries.During advancing it got detached in microcatheter.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 manufacturing date ¿ added h3 device evaluated by mfg ¿updated h3 summary attached - updated d4 expiration date - added d9 product available to stryker ¿ updated d9 returned to manufacturer on ¿updated due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the subject coil was returned with a stryker catheter (no allegation against the catheter), the coil delivery wire was seen to be broken 145cm from the proximal contact, the main coil and the distal part of the delivery wire were not returned for analysis, and the introducer sheath was intact.A functional inspection not carried out as the coil was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported coil in catheter friction could not be confirmed; however, the analysis results are consistent with the reported event.The reported main coil prematurely detached/separated during use was unable to be confirmed during analysis as the distal part of the delivery wire or the coil was not returned.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Addition information received was the device was prepared as per the directions for use, there was no damage noted to the packaging prior to opening the packaging, the device was confirmed to be in good condition prior to use on the patient and continuous flush was set up and maintained throughout the clinical procedure.During analysis, the coil delivery wire was found to be broken 145 cm from the proximal contact and the distal segment of the delivery wire was not returned for device analysis.The main coil was not returned for device analysis.It¿s likely when the reported friction was felt trying to advance the coil during the procedure manipulation and handling of the device would have contributed to the damaged noted to the device during analysis and the reported event.The as analyzed ¿coil delivery wire broken/fractured during use¿ will be assigned handling damage as this complaint appears to be associated with handling of the product or portion of the product during the procedure.The as reported ¿coil in catheter friction¿ will be assigned procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of undeterminable will be assigned to as reported ¿main coil prematurely detected/separated during use¿ as the issue could not be confirmed during analysis as the main coil or the distal segment of the coil delivery wire were not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure, the subject coil could not be advanced to proximal of microcatheter even after several tries.During advancing it got detached in microcatheter.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
|
| |
|
Search Alerts/Recalls
|