C.R. BARD, INC. (BASD) -3006260740 4F SINGLE LUMEN GROSHONG NXT CLEARVUE PICC WITH SHERLOCK TLS STYLET; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer "hole was found in the catheter prior to insertion".
 
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of regs2794 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the customer "hole was found in the catheter prior to insertion".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 4fr s/l groshong catheter.The sample was received with an attached t-lock assembly and inlaid tls stylet.Light usage residues were observed on the sample.Microscopic inspection of the sample did not reveal any evidence of previous or current leakage.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration with no observed leaks.No leaks were observed during pressurization of the sample.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by the customer pinpoint leak at the 42cm depth marking "prior to securing and using it on the patient" a new device was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4F SINGLE LUMEN GROSHONG NXT CLEARVUE PICC WITH SHERLOCK TLS STYLET
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15880913
Report Number3006260740-2022-05527
Device Sequence Number19465898
Product Code LJS
UDI-Device Identifier00801741036026
UDI-Public(01)00801741036026
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K034020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberN/A
Device Catalogue Number9655405
Device Lot NumberREGS2794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer01/25/2023
02/03/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date02/02/2023
02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-