MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 3MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 3MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035443060
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure, the subject coil could not be advanced to distal of microcatheter even after several tries.The operator found the subject coil detached inside the microcatheter during delivery.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned.
 
Event or Problem Description
It was reported that during procedure, the subject coil could not be advanced to distal of microcatheter even after several tries.The operator found the subject coil detached inside the microcatheter during delivery.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Event or Problem Description
It was reported that during procedure, the subject coil could not be advanced to distal of microcatheter even after several tries.The operator found the subject coil detached inside the microcatheter during delivery.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the main coil was seen to be detached/separated, the main coil was seen to be kinked/bent and stretched, the coil delivery wire was seen to be kinked/bent, and the coil introducer sheath was intact.A functional inspection was not carried out due to the damage noted to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information provided by the customer indicated that the coil could not be advanced to distal of the microcatheter and the coil detached inside the microcatheter during delivery.An assignable cause of procedural factors was assigned to the as reported defects: coil in catheter friction; main coil prematurely detached/separated during use and as analyzed defects main coil kinked/bent, main coil stretched, and main coil prematurely detached/separated during use, as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of handling damage was assigned to the as analyzed defects 'coil delivery wire kinked/bent' as the issue is due to handling of the product or portion of the product during the clinical procedure, upon removal of the product from the packaging, or preparation of the product prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 3MM X 6CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15881099
Report Number3008881809-2022-00600
Device Sequence Number5445957
Product Code HCG
UDI-Device Identifier07613252600329
UDI-Public07613252600329
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2025
Device Model NumberM0035443060
Device Catalogue NumberM0035443060
Device Lot Number23665099
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer01/11/2023
01/31/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/12/2023
02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NEUROFORM EZ STENT (STRYKER); SL-10 CATHETER (STRYKER); SYNCHRO GUIDEWIRE (STRYKER); TARGET COILS (STRYKER); XT-27 CATHETER (STRYKER)
Patient SexUnknown
Patient RaceAsian
-
-