| Catalog Number |
07028008190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter questioned positive results for 2 patient samples tested for elecsys toxo igg (toxo igg) on a cobas e 801 analytical unit.On (b)(6) 2022 patient 1 (id (b)(6)) result from the e801 analyzer was 82.1 iu/ml (positive).The result from the vidas method was 6.0 iu/ml (indeterminate).On (b)(6) 2022 patient 2 (id (b)(6)) result from the e801 analyzer was 34.2 iu/ml (positive) the result from the liaison diasorin method was < 3.0 iu/ml (negative).The result from the vidas (predetermination) method was 5.0 iu/ml (indeterminate).No questionable results were reported outside of the laboratory.The e801 analyzer serial number was (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and qc were acceptable.Sample material was requested for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4, expiration date was updated.Both samples were received for investigation and tested with the elecsys toxo igg assay, an in-house neutralization assay, and a competitor toxo igg assay.The samples were also tested for elecsys toxo igg avidity and elecsys toxo igm.For patient 1: the elecsys toxo igg result was 83.8 iu/ml (positive).The sample was neutralizable (10.0% of remaining titer).Recomline blot toxo igg: positive.Elecsys toxo igg avidity: 92% (high avidity).Elecsys toxo igm: 0.222 coi (negative).For patient 2: the elecsys toxo igg result was 34.1 iu/ml (positive).The sample was neutralizable (10.7% of remaining titer).Recomline blot toxo igg: positive.Elecsys toxo igg avidity: 60% (low avidity).Elecsys toxo igm: 0.232 coi (negative).The customer's positive toxo igg results were reproduced during the investigation.Based on the results produced during the investigation, the samples were assessed to be correctly positive.The reagent performed within specification.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|