BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; STYLET FOR CATHETER, GASTRO-UROLOGY Back to Search Results
Model Number M0066401040
Medical Device Problem Codes Break (1069); Unraveled Material (1664); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).One amplatz guidewire was returned.During product analysis, the corewire was broken approximately seven cm from the distal tip and the coil was unraveled.The reported complaint of coil unraveled is confirmed.An investigation to address this problem is in progress.Based on the condition of the returned device, engineers determined that the evidence gathered through investigation determines that the most probable cause is manufacturing deficiency; since there were problems traced to manufacturing process.
 
Event or Problem Description
It was reported to boston scientific corporation that a amplatz was used during a retrograde intrarenal surgery (rirs) procedure performed on (b)(6) 2022.During the procedure, the coil loosens.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation results: the corewire was broken approximately seven cm from the distal tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ SUPER STIFF
Common Device Name
STYLET FOR CATHETER, GASTRO-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15882338
Report Number3005099803-2022-06910
Device Sequence Number2532390
Product Code EZB
UDI-Device Identifier08714729752660
UDI-Public08714729752660
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2024
Device Model NumberM0066401040
Device Catalogue Number640-104S
Device Lot Number0028041277
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2022
Initial Report FDA Received Date11/29/2022
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
-
-