EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
LOGIC CR FEMORAL CEM, LEFT, SZ 4 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t, 3586055, 02-012-51-4009 - logic tib insert impl crc, sz 4, 9mm, 2566348 and 200-05-26 - inset patella 26mm, 2904035.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported a 68 yo male patient, initial left knee implanted on (b)(6) 2014, underwent a revision procedure on (b)(6) 2022, approximately 8 years 4 months post the initial procedure.The patient presented to the surgeon¿s office with consistent complaints of pain, stiffness, and swelling.It had been previously aspirated in the office.The x-rays taken showed possible osteolysis and made the surgeon question if the implants were well fixed within the knee.The poly implant was run in the recall database and did come back as an affected poly within the current knee poly recall.The patient was then scheduled for a left tka revision possible poly swap.The surgeon performed a synovectomy.The surgeon requested and took cultures from the deep joint fluid.It was mentioned that were was obvious osteolysis and tissue inflammation within and around the joint.The surgeon then removed the poly implant and checked the metal components to check if they were well fixed.The poly implant showed significant wear, including pitting and delamination on the articular surface.The surgeon was able to pull the primary femoral implant off the patient's femur bone with his hand.The metal implants were not well fixed.The surgeon then removed all implants (poly, femur, tibia, and patella).The surgeon used a competitor¿s tka system to reimplant a revision tka for this patient.>45 minute surgical delay due to the issues described, as the complete revision caused the patient to be within the or longer than anticipated.No adverse event as a result.The patient was last known to be in stable condition.The implants were asked to be saved, but upon returning to the hospital, the implants that were removed were discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the primary device under consideration has been updated to the logic tib insert impl crc, sz 4, 9mm - product # 02-012-4009 sn (b)(6) - udi (b)(4) - 510k k123342 - manufactured on january 8, 2013 expiration date january 2, 2018.The revision reported may have been the result of loosening, osteolysis, and prosthesis wear.The loosening may have been due to an insufficient bond between the implants and the bone.The prosthesis wear may have been due to third body wear, orientation of the components, patient-related conditions, malalignment between the femoral and tibial components, or any combination of these possibilities.However, this cannot be confirmed as the devices were not available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may have been the result of loosening, osteolysis, and prosthesis wear.The loosening may have been due to an insufficient bond between the implants and the bone.The prosthesis wear may have been due to third body wear, orientation of the components, patient-related conditions, malalignment between the femoral and tibial components, or any combination of these possibilities.However, this cannot be confirmed as the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|