Model Number DFR00V
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an intraocular lens was missing haptic after the iol was inserted into the patient¿s left eye.The lens was removed, and another johnson and johnson lens of the same model and diopter was implanted.There were no medical and/or surgical interventions required, and no patient injury reported.The patient is doing fine post-operation.No further information is available.
 
Additional Manufacturer Narrative
Ethnicity/race: unknown, information asked, but not available.If implanted; give date: not applicable.The iol was removed/replaced in the initial surgery.If explanted; give date: not applicable.The iol was removed/replaced in the initial surgery.The intraocular lens (iol) was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes section d9: returned to manufacturer on: dec 21, 2022 section h3: device evaluated by manufacturer: yes device evaluation: the complaint lens and original handpiece were received loose inside of the original folding carton.The handpiece tray was received as well.Visual inspection under magnification revealed that the complaint lens was received cut in half and with one haptic detached (missing).The lens was cleaned and, no further issues could be identified.Based on the return condition of the lens, no further product evaluation could be identified.The complaint issue of haptic detached was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15883238
Report Number3012236936-2022-02964
Device Sequence Number17467394
Product Code MFK
UDI-Device Identifier05050474750609
UDI-Public(01)05050474750609(17)240907
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer01/03/2023
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight106 KG
-
-