EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number 02-010-01-0260
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.Concomitant medical products: 2216957; 02-012-35-6011 - logic tibia ps mod insrt sz 6 11mm.2600723; 02-012-35-6009 - logic tibia ps mod insrt sz 6 9mm.2892290; 02-012-45-6050 - lgc tibial fit tray cem sz 6f / 5t.3606542; 200-02-41 - three peg patella 41mm.3690508; 200-02-41 - three peg patella 41mm.3817736; 02-012-45-6050 - lgc tibial fit tray cem sz 6f / 5t.Correctional/removal number: z-0021-2022.
 
Event or Problem Description
As reported, the male patient had an initial left tka on (b)(6) 2015.The patient complained of pain and was revised due to a loose femur.The patient was revised to exactech devices.The reported event is unrelated to breakage of a device & did not lead to surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of a combination of prosthesis wear and aseptic loosening between the femoral component and the bone.The loosening may have led to an increase in cement and bone particles in the joint space and resulted in prosthesis wear or prosthesis wear may have contributed to particle-induced osteolysis.However, the contributions of loosening and prosthesis wear to the sequence of events cannot be determined as the devices were not available for evaluation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of femoral loosening and prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral component to the femoral bone, may have lead to loosening at the cement-implant interface; however, this cannot be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 6
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15883277
Report Number1038671-2022-01530
Device Sequence Number4939397
Product Code JWH
UDI-Device Identifier10885862001184
UDI-Public10885862001184
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/30/2016
Device Model Number02-010-01-0260
Device Catalogue Number02-010-01-0260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date11/09/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer08/21/2023
02/04/2025
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date09/04/2023
02/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-