| Model Number |
204-23-11 |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.Concomitant medical products: patella - cn: 200-02-35; sn: (b)(4).Tibia plateau - cn: 200-04-33; sn: (b)(4).Femur - cn: 234-02-03; (b)(4).
|
| |
|
Event or Problem Description
|
|
As reported, the patient had an initial tka in (b)(6) 2018.The patient started to feel pain and discomfort in the knee around (b)(6) 2021 and consulted with his surgeon in (b)(6) 2021 with swelling and limping when walking.The patient underwent complementary exams that indicated loosening of the prosthesis with polyethylene wear and bone lysis in the tibial plateau.The patient was revised with a spacer in late (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of femoral component loosening and wear of the tibial insert.Loosening of the femoral component may have preceded and contributed to wear of the tibial insert and subsequent particle-induced osteolysis or particle-induced osteolysis from instability and wear of the insert may have contributed to component loosening and further damage of the insert.However, the individual contributions of femoral component loosening, prosthesis wear, osteolysis, and infection to the sequence of events cannot be determined as serial radiographs were not available and the device was not received for evaluation.H7: z-0019-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported was likely the result of femoral component loosening and wear of the tibial insert.Loosening of the femoral component may have preceded and contributed to wear of the tibial insert and subsequent particle-induced osteolysis or particle-induced osteolysis from instability and wear of the insert may have contributed to component loosening and further damage of the insert.However, the individual contributions of femoral component loosening, prosthesis wear, osteolysis, and infection to the sequence of events cannot be determined as serial radiographs were not available and the device was not received for evaluation if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|